[Performance, Structure and Components]: This device consists of tube body, connector,
cuff (if any), one-way valve, pilot balloon and inflation tube. [Indications for Use]: This device is indicated for clinical intubation and anesthesia use during emergency treatment. [Instructions for Use]: This device is indicated for clinical intubation and anesthesia use during emergency treatment. 1.Select appropriate size of tracheal tube according to different patients. 2.After opening the package (if any cuff), insert a 10ml syringe into the inflation valve. (Noticed from the pilot balloon that the end of the valve is pushed out by more than 1mm.) Suction air and check for smooth flow and unbroken cuff. 3.When having difficulties in suctioning air, straighten the pilot balloon to make air flow smooth. 4.In accordance with different specifications inflate the cuff (if any) with an amount of air (where fitted), observing the pilot balloon plumping up. 5.After intubation, frequently add in an amount of saline solution to prevent dirt bond and keep the tube unclogged and patient breathing smoothly. 6.Regularly observe that the pilot balloon is inflated normally. 7.Deflate the cuff and easily remove the tube after use.
|
||||||||
Copyright QA Medical Co.,Ltd. SHFDAD No. 20101767 |
2nd Floor, Building 6, 328 Jinglian Rd., Minhang District, Shanghai, P.R.China